Message: Short circuit on the test object 1. The SECUTEST checks whether there is a short circuit between N and L, or between L / N and SL. For the firmware versions V7.08 the limit was 2.5 ohms.
In version V7.08, the limit has been lowered to 1.5 ohms and the measurement time for the short-circuit test has been extended.
2. The SECUTEST performs a leakage current measurement. In the case of test objects with a capacitive input circuit, higher leakage currents can occur here and the SECUTEST reports a short circuit. Under SETUP LIMIT
For the firmware versions LIMIT VALUES, the value "Ground fault at 15.0 mA" can be increased.