PG415: Measurement of SK1 device - protective conductor not accessible

Subject: SECUTEST BASE / BASE10

You are testing a device of protection class I, but the protective conductor is not accessible.

Examples are coffee machines, pumps, power supplies.

Press the button with the wrench.

Press the R-PE button so that it says "off".

Confirm the entry with the button on the green tick.

Now the protective conductor test is omitted.

Note in the log that the protective conductor was not accessible, but the device was otherwise in order.